Simultaneous and interleaved acquisition of NMR signals from different nuclei with a clinical MRI scanner.

Details

Ressource 1Download: 26608834_BIB_BC80D87C2556.pdf (955.09 [Ko])
State: Public
Version: Final published version
License: CC BY 4.0
Serval ID
serval:BIB_BC80D87C2556
Type
Article: article from journal or magazin.
Collection
Publications
Institution
Title
Simultaneous and interleaved acquisition of NMR signals from different nuclei with a clinical MRI scanner.
Journal
Magnetic resonance in medicine
Author(s)
Meyerspeer M., Magill A.W., Kuehne A., Gruetter R., Moser E., Schmid A.I.
ISSN
1522-2594 (Electronic)
ISSN-L
0740-3194
Publication state
Published
Issued date
11/2016
Peer-reviewed
Oui
Volume
76
Number
5
Pages
1636-1641
Language
english
Notes
Publication types: Evaluation Study ; Journal Article
Publication Status: ppublish
Abstract
Modification of a clinical MRI scanner to enable simultaneous or rapid interleaved acquisition of signals from two different nuclei.
A device was developed to modify the local oscillator signal fed to the receive channel(s) of an MRI console. This enables external modification of the frequency at which the receiver is sensitive and rapid switching between different frequencies. Use of the device was demonstrated with interleaved and simultaneous <sup>31</sup> P and <sup>1</sup> H spectroscopic acquisitions, and with interleaved <sup>31</sup> P and <sup>1</sup> H imaging.
Signal amplitudes and signal-to-noise ratios were found to be unchanged for the modified system, compared with data acquired with the MRI system in the standard configuration.
Interleaved and simultaneous <sup>1</sup> H and <sup>31</sup> P signal acquisition was successfully demonstrated with a clinical MRI scanner, with only minor modification of the RF architecture. While demonstrated with <sup>31</sup> P, the modification is applicable to any detectable nucleus without further modification, enabling a wide range of simultaneous and interleaved experiments to be performed within a clinical setting. Magn Reson Med 76:1636-1641, 2016. © 2015 The Authors. Magnetic Resonance in Medicine published by Wiley Periodicals, Inc. on behalf of International Society for Magnetic Resonance in Medicine. This is an open access article under the terms of the Creative Commons Attribution License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
Keywords
Carbon-13 Magnetic Resonance Spectroscopy/instrumentation, Computer-Aided Design, Equipment Design, Equipment Failure Analysis, Magnetic Resonance Imaging/instrumentation, Molecular Imaging/instrumentation, Phosphorus Isotopes/pharmacokinetics, Reproducibility of Results, Sensitivity and Specificity, Systems Integration, MRI, MRS, X-nucleus, interleaved, multinuclear, simultaneous
Pubmed
Web of science
Open Access
Yes
Create date
01/12/2015 9:12
Last modification date
12/01/2022 8:13
Usage data