Cell salvage using the continuous autotransfusion device CATSmart - an observational bicenter technical evaluation

  • Background: The use of cell salvage and autologous blood transfusion has become an important method of blood conservation. So far, there are no clinical data about the performance of the continuous autotransfusion device CATSmart. Methods: In total, 74 patients undergoing either cardiac or orthopedic surgery were included in this prospective, bicenter and observational technical evaluation to validate red cell separation process and washout quality of CATSmart. The target of red cell separation process was defined as a hematocrit value in the packed red cell unit of 55–75% and of washout quality of 80–100% removal ratio. Results: Hematocrit values measured by CATSmart and laboratory analysis were 78.5% [71.3%; 84.0%] and 73.7% [67.5%; 75.5%], respectively. Removal ratios for platelets 94.7% [88.2%; 96.7%], free hemoglobin 89.3% [85.2%; 94.9%], albumin 97.9% [96.6%; 98.5%], heparin 99.9% [99.9%; 100.0%], and potassium 92.5% [90.8%; 95.0%] were within the target range while removal of white blood cells was slightly worse 72.4% [57.9%; 87.3%]. Conclusion: The new autotransfusion device enables sufficient red cell separation and washout quality.

Download full text files

Export metadata

Metadaten
Author:Simone LindauGND, Madeline Kohlhaas, Michael Nosch, Suma ChoorapoikayilGND, Kai ZacharowskiORCiDGND, Patrick MeybohmORCiDGND
URN:urn:nbn:de:hebis:30:3-508306
DOI:https://doi.org/10.1186/s12871-018-0651-0
ISSN:1471-2253
Pubmed Id:https://pubmed.ncbi.nlm.nih.gov/30541447
Parent Title (English):BMC anesthesiology
Publisher:BioMed Central
Place of publication:[s. l.]
Document Type:Article
Language:English
Year of Completion:2018
Date of first Publication:2018/12/12
Publishing Institution:Universitätsbibliothek Johann Christian Senckenberg
Release Date:2019/08/26
Tag:Auto transfusion; Cell salvage; Hematocrit value
Volume:18
Issue:1, Art. 189
Page Number:5
First Page:1
Last Page:5
Note:
Open Access: This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.
HeBIS-PPN:453973574
Institutes:Medizin / Medizin
Dewey Decimal Classification:6 Technik, Medizin, angewandte Wissenschaften / 61 Medizin und Gesundheit / 610 Medizin und Gesundheit
Sammlungen:Universitätspublikationen
Licence (German):License LogoCreative Commons - Namensnennung 4.0