Active Middle Ear Implant Evoked Auditory Brainstem Response Intensity-Latency Characteristics

  • Objective To analyze intensity-latency functions of intraoperative auditory evoked brainstem responses (ABRs) to stimulation by the Vibrant Soundbridge (VSB) active middle ear implant with respect to coupling efficiency, VSB evoked ABR thresholds, and coupling modality [oval window (OW) placement vs. Incus placement and vs. round window (RW) placement]. Study Design Exploratory study. Setting Bi-centric study at tertiary referral centers. Patients Twenty-four patients (10 female, 14 male, mean age: 58 years) who received a VSB. Outcome Measures Wave-V intensity-latency functions of intraoperative VSB evoked ABRs using a modified audio processor programmed to preoperative bone conduction thresholds for stimulation. Threshold level correction to coupling efficiency and ABR thresholds. Individual plots and exponential function fits. Results After ABR threshold level correction, the latency functions could be aligned. A large variance of latencies was observed at individual threshold level. Wave-V latency was longest in the Incus placement subgroup (9.73 ms, SD: 1.04) as compared to OW placement subgroup (9.47 ms, SD: 1.05), with the shortest latency in the RW placement subgroup (8.99 ms, SD: 0.68). For increasing stimulation levels, the variance decreased with intensity-latency function slopes converging toward a steady-state (saturation) latency caused by saturation of audio processor (stimulation) gain. Latency saturation was reached at a stimulation level of 50 dB nHL for the OW placement subgroup, 35 dB nHL for the Incus placement subgroup, and 30 dB nHL for the RW placement subgroup. The latency and saturation results indicated decreased dynamic range for RW placement, i.e., reverse stimulation. Conclusions VSB evoked ABR wave-V intensity-latency function slopes were similar to acoustic stimulation at high stimulation levels with a shift toward longer latencies caused by audio processor signal delay. Saturation of latencies occurred for higher stimulation levels due to saturation of audio processor gain. Thus, the analysis of VSB evoked intensity-latency functions appears to allow for the objective assessment of a patient's individual dynamic range. This can further improve diagnostics as well as intraoperative and postoperative quality control.

Download full text files

Export metadata

Additional Services

Search Google Scholar

Statistics

frontdoor_oas
Metadaten
Author: Laura Fröhlich, Alexander Müller, Miriam H. Kropp, Parwis Mir-Salim, Oliver Dziemba, Tobias Oberhoffner, Stefan K. Plontke, Torsten Rahne
URN:urn:nbn:de:gbv:9-opus-59108
DOI:https://doi.org/10.3389/fneur.2021.739906
ISSN:1664-2295
Parent Title (English):Frontiers in Neurology
Publisher:Frontiers Media S.A.
Place of publication:Lausanne
Document Type:Article
Language:English
Date of first Publication:2022/01/20
Release Date:2022/07/05
Tag:active middle ear implant; auditory brainstem response; coupling efficiency; latency; objective measures
GND Keyword:-
Volume:12
Article Number:739906
Page Number:10
Faculties:Universitätsmedizin / Klinik und Poliklinik für Hals-, Nasen-, Ohrenkrankheiten
Licence (German):License LogoCreative Commons - Namensnennung