Please use this identifier to cite or link to this item: http://dx.doi.org/10.25673/35273
Title: Mittelfristige klinische und radiologische Ergebnisse der Nanos®-Kurzschaftendoprothese
Author(s): Fischer, ChristianLook up in the Integrated Authority File of the German National Library
Referee(s): Wohlrab, DavidLook up in the Integrated Authority File of the German National Library
Hofmann, Gunther Olaf
Gutteck, NataliaLook up in the Integrated Authority File of the German National Library
Granting Institution: Martin-Luther-Universität Halle-Wittenberg
Issue Date: 2020
Extent: 1 Online-Ressource (89 Seiten)
Type: HochschulschriftLook up in the Integrated Authority File of the German National Library
Type: PhDThesis
Exam Date: 2020-12-07
Language: German
URN: urn:nbn:de:gbv:3:4-1981185920-354835
Abstract: Ziel dieser retrospektiven Studie war die Erhebung mittelfristiger klinischer und radiologischer Ergebnisse der Nanos®-Kurzschaftendoprothese. Insgesamt wurden die Daten von insgesamt 51 Patienten (54 Nanos®-Kurzschäften) erhoben. Das durchschnittliche follow-up betrug 97,8 Monate (8,2 Jahre). Evaluiert wurde die subjektive Patientenzufriedenheit bzgl. Schmerz und Funktion nach Kurzschaft-Implantation sowie das objektivierbare klinische Outcome im Harris Hip Score und im Forgotten Joint Score. Anhand radiologischer Untersuchungen wurden Aussagen zu einer postoperativen Veränderung der Beinlänge, der Schaftposition, des femoralen Offsets sowie des horizontalen bzw. vertikalen Drehzentrums getroffen. Zusätzlich wurden das Auftreten einer Angulation bzw. longitudinalen Migration der Kurzschäfte sowie das Auftreten radioluzenter Linien und periartikulärer Ossifikationen untersucht, um Aussagen zur Standzeit sowie zur Rekonstruierbarkeit der patientenindividuellen Anatomie treffen zu können.
The aim of this retrospective study was to obtain medium-term clinical and radiological data on the results of the Nanos®-short-stem endoprosthesis. In this regard, the complete data of 51 (54 Nanos®-short-stems) was reviewed. The average follow-up was 97.8 months (8.2 years). We evaluated patient satisfaction regarding pain and function as well as the postoperative Harris Hip Score and Forgotten Joint Score. Postoperative radiographic evaluation included the measurement of leg length discrepancy, changes of shaft axis, femoral offset and horizontal or vertical center of rotation. Potential postoperative shaft angulation or axial shaft migration was also determined. Radiographic images were checked for radiolucent lines and heterotopic ossification. Statements on the long-term stability of the short-stem prosthesis and the possibility regarding the reconstruction of the patient-specific anatomy were made. We found no cases of aseptic loosening or other causes for revision.
URI: https://opendata.uni-halle.de//handle/1981185920/35483
http://dx.doi.org/10.25673/35273
Open Access: Open access publication
License: In CopyrightIn Copyright
Appears in Collections:Interne-Einreichungen

Files in This Item:
File Description SizeFormat 
Promotion Christian Fischer.pdf4.15 MBAdobe PDFThumbnail
View/Open