An insert with less than spherical medial conformity causes a loss of passive internal rotation after calipered kinematically aligned TKA

Please always quote using this URN: urn:nbn:de:bvb:20-opus-266710
  • Introduction In total knee arthroplasty (TKA), the level of conformity, a medial stabilized (MS) implant, needs to restore native (i.e., healthy) knee kinematics without over-tensioning the flexion space when the surgeon chooses to retain the posterior cruciate ligament (PCL) is unknown. Whether an insert with a medial ball-in-socket conformity and lateral flat surface like the native knee or a less than spherical medial conformity restores higher and closer to native internal tibial rotation without anterior lift-off, an over-tensionIntroduction In total knee arthroplasty (TKA), the level of conformity, a medial stabilized (MS) implant, needs to restore native (i.e., healthy) knee kinematics without over-tensioning the flexion space when the surgeon chooses to retain the posterior cruciate ligament (PCL) is unknown. Whether an insert with a medial ball-in-socket conformity and lateral flat surface like the native knee or a less than spherical medial conformity restores higher and closer to native internal tibial rotation without anterior lift-off, an over-tension indicator, when implanted with calipered kinematic alignment (KA), is unknown. Methods and Materials Two surgeons treated 21 patients with calipered KA and a PCL retaining MS implant. Validated verification checks that restore native tibial compartment forces in passive flexion without release of healthy ligaments were used to select the optimal insert thickness. A goniometer etched onto trial inserts with the ball-in-socket and the less than spherical medial conformity measured the tibial rotation relative to the femoral component at extension and 90° and 120° flexion. The surgeon recorded the incidence of anterior lift-off of the insert. Results The insert with the medial ball-in-socket and lateral flat surface restored more internal tibial rotation than the one with less than spherical medial conformity, with mean values of 19° vs. 17° from extension to 90° flexion (p < 0.01), and 23° vs. 20°-120° flexion (p < 0.002), respectively. There was no anterior lift-off of the insert at 90° and 120° flexion. Conclusion An MS insert with a medial ball-in-socket and lateral flat surface that matches the native knee's spherical conformity restores native tibial internal rotation when implanted with calipered KA and PCL retention without over-tensioning the flexion space.show moreshow less

Download full text files

Export metadata

Additional Services

Share in Twitter Search Google Scholar Statistics
Metadaten
Author: Alexander J. Nedopil, Adithya Shekhar, Stephen M. Howell, Maury L. Hull
URN:urn:nbn:de:bvb:20-opus-266710
Document Type:Journal article
Faculties:Medizinische Fakultät / Lehrstuhl für Orthopädie
Language:English
Parent Title (English):Archives of Orthopaedic and Trauma Surgery
ISSN:1434-3916
Year of Completion:2021
Volume:141
Issue:12
Pagenumber:2287–2294
Source:Archives of Orthopaedic and Trauma Surgery 2021, 141(12): 10.1007/s00402-021-04054-0
DOI:https://doi.org/10.1007/s00402-021-04054-0
Pubmed Id:https://pubmed.ncbi.nlm.nih.gov/34264381
Dewey Decimal Classification:6 Technik, Medizin, angewandte Wissenschaften / 61 Medizin und Gesundheit / 610 Medizin und Gesundheit
Tag:calipered; conforming; insert; kinematic alignment; medial stabilized; rotation; spherical; total knee arthroplasty; total knee replacement
Release Date:2022/06/02
Licence (German):License LogoCC BY: Creative-Commons-Lizenz: Namensnennung 4.0 International