1 Introduction

According to current guidelines, the use of remote management (RM) is recommended for patients with cardiac implantable electronic devices (CIEDs) [1, 2]. Several trials have compared in-person evaluations and RM for the follow-up care of CIED patients and have explored the ability of RM to detect problems early, thereby improving patient outcomes [3,4,5,6]. The advent of wireless RM and novel diagnostic features that enable devices to monitor their own functions, record arrhythmias and physiological parameters, and automatically communicate this information to healthcare providers without the active participation of the patient has been critical to achieving the follow-up goals of patient adherence to structured follow-up protocols and to improving the workflow efficiency of device clinics [2, 7, 8]. Successful transmission of RM data by the patient to the healthcare provider relies on the enrollment of the patient in the specific RM system and the subsequent activation and maintenance of RM by the patient. Enrollment in RM has been shown to depend largely upon the local practice of the institution. However, RM activation and transmission by the patient depend upon patient factors [9]. The continuity of monitoring is crucial. Indeed, patients who consistently transmit data by means of RM are at substantially lower risk of death and readmission to hospital [10].

The MyLATITUDE™ Patient App (Boston Scientific) has been developed to encourage patients’ compliance with RM by providing them with information on communicator setup and troubleshooting, connection status of the communicator, scheduled transmissions, and status of the battery of the implanted device.

In the present study, we evaluated the first experience of the adoption of the MyLATITUDE™ Patient App in clinical practice.

2 Methods

From May to July 2021, at 14 Italian arrhythmia centers, patients with a compatible Boston Scientific CIED with RM capabilities were invited to download and install the App on a mobile device.

The use of the App was proposed to all consecutive patients after CIED implantation or, during a scheduled follow-up visit, to consecutive patients already enrolled and monitored at home through the LATITUDE™ platform. The study design was approved by the Institutional Committee on Human Research of each center and informed consent was obtained from all patients. According to individual patient preference, patients were assigned to the App group or the control group.

In the App group, patients received instructions on how to download the App from the App Store of their own mobile device. The LATITUDE™ communicator was given to all patients before discharge or directly delivered to their homes, according to the standard practice of the center. Patients were invited to follow the indications provided by the App in order to activate the communicator and perform the first transmission. If they had any difficulties, patients could contact the center or the Boston Scientific customer support service. After activation, patients were instructed to follow the indications provided by the App in order to solve possible connection problems in the event of notification of “Not Monitored” status or missed scheduled remote transmission. The group of patients already monitored at home through the LATITUDE™ platform and invited to use the App were included in the analysis of the compliance with remote monitoring. An additional group of 60 consecutive patients served as control group and received standard in-clinic training on the activation and use of the communicator, as well as the patients who did not install the App. In November, the monitoring status of all patients with a LATITUDE™ communicator was checked. Patients in “Monitored” status were considered compliant with remote monitoring at medium-term follow-up. All patients were asked to reply to an ad hoc questionnaire to collect anonymous data and evaluate their experience. All data were collected under local standard-of-care conditions of use. The operators at the centers were also asked to complete a questionnaire to provide their feedback on the adoption of the App.

2.1 MyLATITUDE™ Patient App

The App enables patients to independently obtain instructions and information on their communicator and CIED. It works with all Boston Scientific CIEDs that are monitored on the LATITUDE™ NXT platform and can be used by patients and/or their caregivers (Fig. 1). The App is available for iOS and Android mobile phones; it does not connect directly to the patient’s device but receives the data from the server. The App provides easy-to-follow steps that enable patients to set up their communicator. If the patient’s communicator is not in a “Monitored” state, the App notifies the patient and provides resources to troubleshoot the problem, in order to encourage patient compliance with RM and to reduce the center’s burden of managing “Not Monitored” patients. Moreover, the App informs patients of the status of scheduled remote transmission and of the CIED battery.

Fig. 1
figure 1

The MyLATITUDE™ Patient App enables patients to independently obtain instruction and information on their LATITUDE™ NXT Communicator and implanted device

2.2 Statistical analysis

Descriptive statistics are reported as means ± standard deviation for normally distributed continuous variables or medians with 25th to 75th percentiles in the case of skewed distribution. Categorical data were expressed as percentages. Differences between mean data were compared by means of a t-test for Gaussian variables and Mann-Whitney non-parametric test for non-Gaussian variables. Differences in proportions were compared by means of a chi-square analysis. A p-value <0.05 was considered significant for all tests. All statistical analyses were performed by means of R: a language and environment for statistical computing (R Foundation for Statistical Computing, Vienna, Austria).

3 Results

3.1 Participating centers and device management service

The MyLATITUDE™ Patient App was proposed to patients with a compatible CIED at 14 Italian arrhythmia centers. At 11 (79%) centers, the professionals routinely involved in remote device management were nurses and physicians, in accordance with a “Primary Nursing” model. In the remaining 3 (21%) centers, they were physicians alone. The communicator was delivered by nurses at 11 (79%) centers and by physicians at 3 (21%) centers. At 6 (43%) centers, home delivery of the communicator was allowed if preferred by the patient. The professionals involved in training patients in the activation and use of the communicator were nurses at 9 (64%) centers and physicians at 5 (36%) centers.

3.2 Acceptance and installation of the App

The App was proposed to 242 patients: 81 consecutive patients immediately after CIED implantation and 161 patients who were already being monitored at home through the LATITUDE™ platform and who were consecutively enrolled at the time of an in-clinic visit. The App was successfully installed by 177 (73%) patients; 49 (20%) declined, 12 (5%) did not have a suitable mobile phone, and 4 (2%) experienced technical problems that precluded LATITUDE™ communicator activation (lack of landline and mobile telephone coverage). Baseline parameters of the study population are reported in Table 1. Patients who installed the App were younger and had a higher level of education. The App was more frequently accepted by patients already using RM.

Table 1 Baseline parameters of the patients invited to use the App and comparison between those who installed it and those who did not

The App was installed by 127 (72%) patients, by 39 (22%) caregivers, and by patient and caregiver in 11 (6%) cases. Installation problems that did not preclude final activation were reported by 15 (8%) patients. In 12 cases, troubleshooting required telephone support.

3.3 Installation and activation of the communicator

Forty-eight patients successfully followed the indications provided by the App on how to activate the communicator and perform the first transmission. Standard post-implantation in-clinic training was received by the 33 new RM activation patients who did not install the App and by another 60 consecutive patients. Of these latter, one (2%) patient experienced technical problems that precluded communicator activation. The duration of in-clinic training was < 30 min for all patients. The App-guided installation group and the control groups (standard in-clinic training group and App-decliners) were similar in terms of the time required for activation and the need for additional support (Table 2). Patient survey questions on the installation and activation procedure in the groups are reported in Table 2. The level of satisfaction with the information received from the App during the installation procedure was comparable to that provided by in-person training.

Table 2 Details of the installation and activation procedure, and survey questions on the patient experience. Comparison between patients who used the App to install and activate the communicator and those who underwent standard in-clinic training (directly or after declining to use the App)

3.4 Use of the App during follow-up

Of the 177 patients who installed the App, 20 (11%) received notifications of lack of connection during follow-up and 22 (12%) information of missed scheduled transmissions. The median time between notification of “Not monitored” status and resolution was 2 days (<1 week in 90%). After 3 months, 175 (99%) of these 177 communicators were in “Monitored” status. Of the 120 patients without the App installed, 113 (94%) were in “Monitored” status at the end of the observation (p=0.033). In particular, 48 (100%) of 48 patients who installed the App immediately after CIED implantation and 54 (92%) of 59 patients of the standard in-clinic training group (excluding App-decliners) were in “Monitored” status at the end of the observation (p=0.063).

The continuous monitoring of the connection status during follow-up made patients feel reassured (Table 3). Survey questions on operator experience with the App are reported in Table 4. Most operators judged the App to be an effective and efficient tool. Nonetheless, in the opinion of those operators who were not inclined to extensively suggest the use of the App, the best targets are younger and more highly educated patients or those who can be supported by a caregiver.

Table 3 Patient survey questions on the experience with the App during follow-up
Table 4 Survey questions on operator experience with the App (14 respondents)

4 Discussion

In the present study, we described the introduction into clinical practice of a mobile App designed to help CIED patients install and use an RM communicator and to provide notifications on the connection status of the device.

The MyLATITUDE™ Patient App was accepted by a high proportion of CIED patients. Moreover, installation of the App was completed successfully in most patients. The App was more frequently installed by younger and more highly educated patients and especially by those who were already enrolled and monitored at home through the RM platform. Indeed, the implantation of a CIED constitutes a major change for the patient, who might need some time before trying to learn about RM and additional tools. Although in most cases the App was directly installed by the patients themselves, when a caregiver was available, it was installed by this latter. Indeed, the support of a caregiver can help to overcome the reluctance of older and less tech-savvy patients.

Patients who installed the LATITUDE™ communicator under the guidance of the App did so just as successfully as those who had undergone standard post-implantation training of about 30 min in hospital; they had no additional problems, and installation time was similar. Moreover, the level of satisfaction with the App-guided installation was high, and, although approximately 30% of patients responded that they would have preferred in-person training, it is noteworthy that 20% of patients in the control group would have preferred more extensive training. The use of the App to activate the communicators is mainly intended as a potential advantage for the centers, which save the time usually devoted to in-person training and hopefully receive fewer requests for support, as patients can troubleshoot by using the App.

In patients who enabled the App, we recorded notifications of missed transmissions or lack of connection during follow-up. After notification, problems of connection were promptly solved by the patients. Consequently, these patients were able to remain connected with greater continuity than a similar group of patients who did not use the App, with a very high proportion of communicators regularly transmitting data at 3 months (99% versus 94%, p=0.033). This finding was obtained in the overall group of patients who installed the App and confirmed, although with no statistical significance (100% versus 92%, p=0.063), comparing patients who installed the App immediately after CIED implantation and patients of the standard in-clinic training group. This finding may have clinical implications. Indeed, previous analyses revealed not only better survival of patients enrolled in RM programs [11], but also an association between RM and a lower risk of mortality and re-hospitalization in the real world. Specifically, one analysis showed that the adjusted hazard of mortality up to 3 years was significantly lower among patients on active RM at 90 days (HR 0.80, 95% CI 0.76–0.84) [10].

Barriers to the widespread use of RM can arise at the time of program implementation in the center and also owing to difficulties in activating RM and keeping patients monitored. While the implementation depends largely on the local practice environment and the institution, RM activation and transmission are more dependent on patient factors [9]. The adoption of the MyLATITUDE™ Patient App made patients feel reassured, seemed effective in helping them stay compliant with RM, and may possibly promote their engagement in their own healthcare.

4.1 Improving effectiveness and efficiency of remote management of CIEDs

The remote management of CIEDs has been proposed as a strategy to improve the efficiency of device follow-up by replacing in-office device follow-up visits with remote transmissions [12,13,14]. Moreover, RM enables continuous assessment of device-related and clinical parameters, allowing early detection of device malfunction [15,16,17] or clinically relevant events [18,19,20], thereby potentially improving clinical outcomes [11, 21, 22]. Thus, current guidelines suggest the routine use of RM [1, 2], a recommendation that is even stronger in the era of the coronavirus disease 19 (COVID-19) pandemic, in order to maintain a high degree of safety and limit face-to-face interactions [23,24,25,26]. According to a survey by the European Heart Rhythm Association, the main barriers to the adoption of remote monitoring are the lack of reimbursement and the increased workload [27]. Although the COVID-19 pandemic has caused an acceleration in the use of RM of CIEDs, and also the establishment of reimbursement codes in some European regions, the affordability of the model for the centers that decide to implement an RM service remains a critical issue [28, 29]. The adoption of tools like the MyLATITUDE™ Patient App may be an efficient way to ensure continuity of patient monitoring while maintaining under control the workload at the centers. This could be achieved by reducing the time spent on training and support for technical troubleshooting and on managing missed transmissions, which are known to impair the clinical efficiency of RM [30].

We recently implemented direct home delivery of the communicator and remote training of patients during the COVID-19 lockdown period in Italy [31]. This strategy proved feasible, enabling the RM of all previously unmonitored CIED patients, without requiring access to the hospital [32]. The MyLATITUDE™ Patient App could be effectively adopted to support that strategy, by enabling patients to activate their communicator even without traditional in-person training. As proposed by many RM programs, after CIED implantation, the implementation of RM could be deferred [2]. When patients are at home and have had the time to recover, accept the implanted device, and process the first CIED-related information, they may receive the communicator and proceed with installation by themselves, with the aid of the mobile App.

4.2 Limitations

Our findings have potential limitations. This was an observational non-randomized study. Some analyses were performed with patients self-selecting whether to be in the intervention or control group. Thus, a bias could affect our findings. As the project was limited to a single RM platform, our results may not be applicable to other systems. Moreover, we cannot exclude possible differences in the implementation of the initiative among centers, with an impact on the degree of success.

5 Conclusion

The MyLATITUDE™ Patient App proved to be well accepted by CIED patients. It offered support for communicator installation and management and reassured patients. Its use allowed patients to remain connected with greater continuity during follow-up.