Endoluminal surgical triangulation: overcoming challenges of colonic endoscopic submucosal dissections using a novel flexible endoscopic surgical platform: feasibility study in a porcine model.

Details

Ressource 1Download: REF.pdf (411.94 [Ko])
State: Public
Version: Final published version
License: Not specified
It was possible to publish this article open access thanks to a Swiss National Licence with the publisher.
Serval ID
serval:BIB_ED16630D20EE
Type
Article: article from journal or magazin.
Collection
Publications
Institution
Title
Endoluminal surgical triangulation: overcoming challenges of colonic endoscopic submucosal dissections using a novel flexible endoscopic surgical platform: feasibility study in a porcine model.
Journal
Surgical Endoscopy
Author(s)
Diana M., Chung H., Liu K.H., Dallemagne B., Demartines N., Mutter D., Marescaux J.
ISSN
1432-2218 (Electronic)
ISSN-L
0930-2794
Publication state
Published
Issued date
2013
Volume
27
Number
11
Pages
4130-4135
Language
english
Notes
Publication types: Journal Article ; Research Support, Non-U.S. Gov'tPublication Status: ppublish
Abstract
BACKGROUND: Colonic endoscopic submucosal dissection (ESD) is challenging as a result of the limited ability of conventional endoscopic instruments to achieve traction and exposure. The aim of this study was to evaluate the feasibility of colonic ESD in a porcine model using a novel endoscopic surgical platform, the Anubiscope (Karl Storz, Tüttlingen, Germany), equipped with two working channels for surgical instruments with four degrees of freedom offering surgical triangulation.
METHODS: Nine ESDs were performed by a surgeon without any ESD experience in three swine, at 25, 15, and 10 cm above the anal verge with the Anubiscope. Sixteen ESDs were performed by an experienced endoscopist in five swine using conventional endoscopic instruments. Major ESD steps included the following for both groups: scoring the area, submucosal injection of glycerol, precut, and submucosal dissection. Outcomes measured were as follows: dissection time and speed, specimen size, en bloc dissection, and complications.
RESULTS: No perforations occurred in the Anubis group, while there were eight perforations (50 %) in the conventional group (p = 0.02). Complete and en bloc dissections were achieved in all cases in the Anubis group. Mean dissection time for completed cases was statistically significantly shorter in the Anubis group (32.3 ± 16.1 vs. 55.87 ± 7.66 min; p = 0.0019). Mean specimen size was higher in the conventional group (1321 ± 230 vs. 927.77 ± 229.96 mm(2); p = 0.003), but mean dissection speed was similar (35.95 ± 18.93 vs. 23.98 ± 5.02 mm(2)/min in the Anubis and conventional groups, respectively; p = 0.1).
CONCLUSIONS: Colonic ESDs were feasible in pig models with the Anubiscope. This surgical endoscopic platform is promising for endoluminal surgical procedures such as ESD, as it is user-friendly, effective, and safe.
Pubmed
Web of science
Open Access
Yes
Create date
16/11/2013 14:35
Last modification date
14/02/2022 7:57
Usage data