Superior diagnostic performance of perfusion-cardiovascular magnetic resonance versus SPECT to detect coronary artery disease: The secondary endpoints of the multicenter multivendor MR-IMPACT II (Magnetic Resonance Imaging for Myocardial Perfusion Assessment in Coronary Artery Disease Trial).

Details

Ressource 1Download: BIB_F2F14DFC5F25.P001.pdf (313.48 [Ko])
State: Public
Version: author
Serval ID
serval:BIB_F2F14DFC5F25
Type
Article: article from journal or magazin.
Collection
Publications
Institution
Title
Superior diagnostic performance of perfusion-cardiovascular magnetic resonance versus SPECT to detect coronary artery disease: The secondary endpoints of the multicenter multivendor MR-IMPACT II (Magnetic Resonance Imaging for Myocardial Perfusion Assessment in Coronary Artery Disease Trial).
Journal
Journal of Cardiovascular Magnetic Resonance
Author(s)
Schwitter J., Wacker C.M., Wilke N., Al-Saadi N., Sauer E., Huettle K., Schönberg S.O., Debl K., Strohm O., Ahlstrom H., Dill T., Hoebel N., Simor T.
Working group(s)
for the MR-IMPACT investigators
ISSN
1532-429X (Electronic)
ISSN-L
1097-6647
Publication state
Published
Issued date
2012
Peer-reviewed
Oui
Volume
14
Number
1
Pages
61
Language
english
Notes
Publication types: JOURNAL ARTICLE
Abstract
ABSTRACT: BACKGROUND: Perfusion-cardiovascular magnetic resonance (CMR) is generally accepted as an alternative to SPECT to assess myocardial ischemia non-invasively. However its performance vs gated-SPECT and in sub-populations is not fully established. The goal was to compare in a multicenter setting the diagnostic performance of perfusion-CMR and gated-SPECT for the detection of CAD in various populations using conventional x-ray coronary angiography (CXA) as the standard of reference. METHODS: In 33 centers (in US and Europe) 533 patients, eligible for CXA or SPECT, were enrolled in this multivendor trial. SPECT and CXA were performed within 4 weeks before or after CMR in all patients. Prevalence of CAD in the sample was 49% and 515 patients received MR contrast medium. Drop-out rates for CMR and SPECT were 5.6% and 3.7%, respectively (ns). The study was powered for the primary endpoint of non-inferiority of CMR vs SPECT for both, sensitivity and specificity for the detection of CAD (using a single-threshold reading), the results for the primary endpoint were reported elsewhere. In this article secondary endpoints are presented, i.e. the diagnostic performance of CMR versus SPECT in subpopulations such as multi-vessel disease (MVD), in men, in women, and in patients without prior myocardial infarction (MI). For diagnostic performance assessment the area under the receiver-operator-characteristics-curve (AUC) was calculated. Readers were blinded versus clinical data, CXA, and imaging results. RESULTS: The diagnostic performance (= area under ROC = AUC) of CMR was superior to SPECT (p = 0.0004, n = 425) and to gated-SPECT (p = 0.018, n = 253). CMR performed better than SPECT in MVD (p = 0.003 vs all SPECT, p = 0.04 vs gated-SPECT), in men (p = 0.004, n = 313) and in women (p = 0.03, n = 112) as well as in the non-infarct patients (p = 0.005, n = 186 in 1-3 vessel disease and p = 0.015, n = 140 in MVD). CONCLUSION: In this large multicenter, multivendor study the diagnostic performance of perfusion-CMR to detect CAD was superior to perfusion SPECT in the entire population and in sub-groups. Perfusion-CMR can be recommended as an alternative for SPECT imaging. TRIAL REGISTRATION: ClinicalTrials.gov, Identifier: NCT00977093.
Pubmed
Web of science
Open Access
Yes
Create date
26/10/2012 19:18
Last modification date
20/08/2019 17:20
Usage data